The IQ-O is a simple and robust quenching system for heat treatment under a controlled atmosphere. The system is based on a quartz glass tube with a water-cooled section and enables flexible, manual quenching processes, even while the system is in operation.
| Temp. máx. 1100 °C, 1200 °C | Horno Ø 120 mm |
| Longitudes de calentamiento 500 mm | Número de zonas 3 zones |
The sample is located inside a quartz glass tube divided into a heated zone and a cooled section.
The sample is moved manually using a mechanical guide through the open gas outlet. This allows the sample to be moved along the axis of the quartz glass tube between the heating and cooling zones and positioned precisely.
This arrangement enables the sample to be moved repeatedly between both zones. Multiple heating and quenching cycles can therefore be carried out under defined conditions without removing the sample from the system.
Cascaded thermocouples
| Exterior dimensions | HxWxD | 1900 x 4000 x 1400 mm | Height of customer’s exhaust system with 100m 3 /h suction capacity |
2300 mm |
| Transport weight | Complete system | 1400 kg |
| Effective space | Inner diameter APM tube | 90 mm | Useful space HxWxD | 50 x 50 x 350 mm |
| Thermal specification | T max (vacuum) | 1200 °C | Tmax (atmospheric pressure) | 1200 °C | Number of heating zones | Up to 3 | max. heating rate (up to 1200°C) | tbd K/min | Cooling down time empty furnace without fast cooling | 7-9 h |
| Electrical connections | Power | 12 kW | Voltage | 3x400 V | Current | 3x28 A | Customer’s pre fuse | 3x35 A |
| Gas supply | Inert gas or Reactive gas | Ar, N2, H2 other gases on request | Flow rate inert or reactive gases | 5-100 l/h | Flow rate inert gas with bypass | 2000 l/h | Consumption compressed air for after burner | 1300 l/h |
| Operation pressure: (furnace clean, cold, dry and empty) | With a double stage rotary pump or turbomolekularpump |
Up to 5x10-5 mbar | Leakage rate | <5x10-3 mbar l/s |
| IQ-S | IQ-L | IQ-M | IQ-O | Usable Volume [mm³] | 50x50x350 | iD50x 200 D50 x 500 50 x 250 |
D 100 | Usable Volume [L] | 7 l | 8 l | 17 l (Medienplan) | Maximum temperature [°C] of oven/sample | 1200 | 1700 / 1600 | 1350 / 1200 | 1300 | Number of heating zones | 1 | 1/3 | 1/3 | Max. outer tube diameter [mm] | 80 | 110 mm / inside 99 | Heating element | MoSi2 | CrAlFe | CrAlFe (APM) | Thermocouple | Type S | Type B | Type S | Type K | Tube material | Ceramics | Inconel / Quartz APM | Quartz glass | Heated Length [mm] | 500 | HTRV 250 or 500mm FHA 200 or 500mm |
200 / 500 | Homogeneous zone +/- 5 K [mm] | 75 | 300 | External Dimensions H x W x D [mm³] | 1900 x 4000 x 1400 | 2100 x 720 x 700 2000 x 1200 x 800 | 1700 x 2500 x 1600 | - | Mechanism | Automatic movement of the sample | Manual movement of the sample | Automatic Movement of the Oven | Manual movement of the sample |