The IQ-M is a specialised quenching system in which the sample remains stationary throughout the entire process. Quenching is achieved by moving the furnace away from the sample, making this solution particularly suitable for sensitive or safety-critical applications.
| Max temp 1100 °C, 1200 °C, 1300 °C | Furnace Ø 50, 60, 80, 110 mm |
| Heated length(s) 250,500,750 mm | Number of heated zones 1,2 ,3 zones |
The sample remains in a fixed position inside the work tube throughout the entire process.
A key feature of the system is the ability to move the furnace away from the sample automatically at a user-defined point in time. This can take place immediately after heating, after a defined holding time at the maximum temperature, or after cooling to a specified target temperature.
As a result, the sample is rapidly transferred from the hot zone into a cooler environment. This enables very fast cooling, even under controlled conditions such as high vacuum.
The furnace is moved automatically by an electric drive integrated directly into the control system, ensuring precise positioning between the heating and quenching zones.
| Exterior dimensions | HxWxD | 1900 x 4000 x 1400 mm | Height of customer’s exhaust system with 100m 3 /h suction capacity |
2300 mm |
| Transport weight | Complete system | 1400 kg |
| Effective space | Inner diameter APM tube | 90 mm | Useful space HxWxD | 50 x 50 x 350 mm |
| Thermal specification | T max (vacuum) | 1200 °C | Tmax (atmospheric pressure) | 1200 °C | Number of heating zones | Up to 3 | max. heating rate (up to 1200°C) | tbd K/min | Cooling down time empty furnace without fast cooling | 7-9 h |
| Electrical connections | Power | 12 kW | Voltage | 3x400 V | Current | 3x28 A | Customer’s pre fuse | 3x35 A |
| Gas supply | Inert gas or Reactive gas | Ar, N2, H2 other gases on request | Flow rate inert or reactive gases | 5-100 l/h | Flow rate inert gas with bypass | 2000 l/h | Consumption compressed air for after burner | 1300 l/h |
| Operation pressure: (furnace clean, cold, dry and empty) | With a double stage rotary pump or turbomolekularpump |
Up to 5x10-5 mbar | Leakage rate | <5x10-3 mbar l/s |
| IQ-S | IQ-L | IQ-M | IQ-O | Usable Volume [mm³] | 50x50x350 | iD50x 200 D50 x 500 50 x 250 |
D 100 | Usable Volume [L] | 7 l | 8 l | 17 l (Medienplan) | Maximum temperature [°C] of oven/sample | 1200 | 1700 / 1600 | 1350 / 1200 | 1300 | Number of heating zones | 1 | 1/3 | 1/3 | Max. outer tube diameter [mm] | 80 | 110 mm / inside 99 | Heating element | MoSi2 | CrAlFe | CrAlFe (APM) | Thermocouple | Type S | Type B | Type S | Type K | Tube material | Ceramics | Inconel / Quartz APM | Quartz glass | Heated Length [mm] | 500 | HTRV 250 or 500mm FHA 200 or 500mm |
200 / 500 | Homogeneous zone +/- 5 K [mm] | 75 | 300 | External Dimensions H x W x D [mm³] | 1900 x 4000 x 1400 | 2100 x 720 x 700 2000 x 1200 x 800 | 1700 x 2500 x 1600 | - | Mechanism | Automatic movement of the sample | Manual movement of the sample | Automatic Movement of the Oven | Manual movement of the sample |