Each package comprises a work tube of the specified diameter with insulation plugs or heat shields appropriate for the operating atmosphere. A pair of end seals is included in the gas atmosphere and vacuum packages.
A 3 mm diameter probe thermocouple is included in all work tube packages for furnaces up to 1200 °C, for connection to a separate display or recorder. Above 1200 °C one end seal will have a 10 mm diameter thermocouple gland with blanking plug to suit a probe thermocouple (probe thermocouple must be ordered separately).
The work tube material in the packages are:
- For TF, TS & TG 1200°C furnaces there is a choice of mullite C610 / IAP or quartz.
- For TF 1600°C furnaces there is a choice of RCA or mullite C610 / IAP.
- For F-range 1350°C furnaces there is a choice of mullite or APM.
- For HTRH & HTRV 1800°C range furnaces RCA tubes are used.
Work tube package 1 - for air
Tube furnace work tube package for air contains:
- Standard length work tube
- 2 x insulation plugs for standard length work tube
- Support bracket for insulation plugs & thermocouple
- Probe thermocouple access:
- Up to 1200 °C: Probe thermocouple gland complete with type N thermocouple
- Above 1200 °C: 10 mm gland with PTFE seal and blanking plug to suit optional type R or B or 20/40 probe thermocouple assembly
Avaliable for all Tube Furnaces
Work tube package 2 - for gas atmosphere
The work tube package for gas atmosphere is suitable for either inert gas or hydrogen atmosphere.
If used with hydrogen a gas safety system MUST be used.
Tube furnace work tube package for gas atmosphere contains:
- Extended length work tube
- 2 x insulation plugs for extended length work tube
- Work tube end seals:
- 1 x gas inlet/outlet pipe
- 1 x gas inlet/outlet pipe + thermocouple gland
- Probe thermocouple access:
- Up to 1200 °C: Probe thermocouple gland complete with type N thermocouple
- Above 1200 °C: 10 mm gland with PTFE seal and blanking plug to suit optional type R or B or 20/40 probe thermocouple assembly
Avaliable for all Tube Furnaces
Work tube package 3 - for vacuum atmosphere
Available for work tube outer diameters 32, 60, 86, 125, 150, 200 mm. Suitable for vacuum pressure to 5 x 10-5 mbar depending on type of vacuum pump used.
Tube furnace work tube package for vacuum atmosphere contains:
- Extended length work tube
- 2 x radiation shields
- Work tube end seals:
- 1 x NW16 vacuum flange + thermocouple gland
- 2 x end plates: 1 x NW25 (compatible with low vacuum pump package); 1 x NW40 for 60 mm work tube outer diameter or 1 x ISO-K 63 for 86, 125, 150 & 200 mm work tube outer diameters (compatible with high vacuum pump package)
- Probe thermocouple access:
- Up to 1200 °C:
Probe thermocouple gland complete with type N thermocouple - Above 1200 °C:
10 mm gland with PTFE seal and blanking plug to suit optional type R or B or 20/40 probe thermocouple assembly
- Up to 1200 °C:
Avaliable for all Tube Furnaces
WORK TUBE PACKAGE 4 VACUUM APPLICATIONS WITH LOW LEAKAGE AND INERT GAS ATMOSPHERE
The package also includes furnace mounted flange holders. The vertical versions include a stand and differ slightly.
Tube furnace work tube package for vacuum atmosphere contains:
- Overflow valve
- Water cooled flange requires 2-4 l/min cooling water
- Fibre wool blankets
- Radiation shields
- Gas outlet with hand valve
- Extended length work tube
- Gas inlet / 6 mm compression fitting
- Flow controller mounted in the frame