The IQ-L is a vertically designed quenching system in which the sample is quenched by lowering it into a quenching medium.
| Maximaltemperatur 1300 °C, 1600 °C, 1700 °C, 1800 °C | Ofen Ø 50, 60, 70, 80 mm |
| Beheizte Länge(n) 250,500,750 mm | Anzahl der Heizzonen 1,2 ,3 zones |
1. Positioning and Heating
The sample is positioned centrally inside the ceramic work tube using a wire attached to a hook. In this initial position, the heat-treatment process begins under a defined atmosphere.
2. Separation of the Heating and Quenching Zones
The quench tank is located below the heating zone and is separated from it by a manually operated valve.
3. Activation of the Quenching Mechanism
Once the heat treatment is complete, the valve is opened to provide access to the quenching zone.
The suspension wire is attached to a hook by means of an eyelet. The hook is connected to a rotating mechanism. Manually rotating the hook by 180° releases the eyelet, allowing the sample and suspension wire to drop into the quenching zone.
4. Quenching in the Quench Tank
The sample falls directly into the quenching zone, where it is rapidly cooled in a defined, customer-specific medium.
| Exterior dimensions | HxWxD | 1900 x 4000 x 1400 mm | Height of customer’s exhaust system with 100m 3 /h suction capacity |
2300 mm |
| Transport weight | Complete system | 1400 kg |
| Effective space | Inner diameter APM tube | 90 mm | Useful space HxWxD | 50 x 50 x 350 mm |
| Thermal specification | T max (vacuum) | 1200 °C | Tmax (atmospheric pressure) | 1200 °C | Number of heating zones | Up to 3 | max. heating rate (up to 1200°C) | tbd K/min | Cooling down time empty furnace without fast cooling | 7-9 h |
| Electrical connections | Power | 12 kW | Voltage | 3x400 V | Current | 3x28 A | Customer’s pre fuse | 3x35 A |
| Gas supply | Inert gas or Reactive gas | Ar, N2, H2 other gases on request | Flow rate inert or reactive gases | 5-100 l/h | Flow rate inert gas with bypass | 2000 l/h | Consumption compressed air for after burner | 1300 l/h |
| Operation pressure: (furnace clean, cold, dry and empty) | With a double stage rotary pump or turbomolekularpump |
Up to 5x10-5 mbar | Leakage rate | <5x10-3 mbar l/s |
| IQ-S | IQ-L | IQ-M | IQ-O | Usable Volume [mm³] | 50x50x350 | iD50x 200 D50 x 500 50 x 250 |
D 100 | Usable Volume [L] | 7 l | 8 l | 17 l (Medienplan) | Maximum temperature [°C] of oven/sample | 1200 | 1700 / 1600 | 1350 / 1200 | 1300 | Number of heating zones | 1 | 1/3 | 1/3 | Max. outer tube diameter [mm] | 80 | 110 mm / inside 99 | Heating element | MoSi2 | CrAlFe | CrAlFe (APM) | Thermocouple | Type S | Type B | Type S | Type K | Tube material | Ceramics | Inconel / Quartz APM | Quartz glass | Heated Length [mm] | 500 | HTRV 250 or 500mm FHA 200 or 500mm |
200 / 500 | Homogeneous zone +/- 5 K [mm] | 75 | 300 | External Dimensions H x W x D [mm³] | 1900 x 4000 x 1400 | 2100 x 720 x 700 2000 x 1200 x 800 | 1700 x 2500 x 1600 | - | Mechanism | Automatic movement of the sample | Manual movement of the sample | Automatic Movement of the Oven | Manual movement of the sample |